262
Keysight InfiniiVision 2000 X-Series Oscilloscopes User's Guide
18
Save/Recall (Setups, Screens, Data)
• When saving to a CSV format file, the maximum number of data points is
50,000.
When necessary, the Length control performs a "1 of n" decimation of the data .
For example: if the
Length
is set to 1000, and you are displaying a record that is
5000 data points in length, four of each five data points will be decimated,
creating an output file 1000 data points in length.
When saving waveform data, the save times depend on the chosen format:
See Also
•
•
•
"Minimum and Maximum Values in CSV Files"
To save Lister data files
Lister data files can be saved to an external USB storage device.
1
Press
[Save/Recall] > Save > Format
; then, turn the Entry knob to select
Lister data
file
.
2
Press the softkey in the second position and use the Entry knob to navigate to
the save location. See
"To navigate storage locations"
3
Press the
Settings
softkey.
In the File Settings Menu, you have these softkeys and options:
•
Setup Info
— when enabled, setup information (vertical, horizontal, trigger,
acquisition, math, and display settings) is also saved in a separate file with a
TXT extension.
4
Finally, press the
Press to Save
softkey.
A message indicating whether the save was successful is displayed.
Data File Format
Save Times
BIN
fastest
ASCII XY
medium
CSV
slowest
Summary of Contents for DSO-X 2014A
Page 1: ...Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide...
Page 8: ...8 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide...
Page 60: ...60 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Page 68: ...68 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 3 Vertical Controls...
Page 118: ...118 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 6 Digital Channels...
Page 124: ...124 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 7 Serial Decode...
Page 130: ...130 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 8 Display Settings...
Page 168: ...168 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 10 Triggers...
Page 176: ...176 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling...
Page 202: ...202 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 13 Cursors...
Page 226: ...226 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 14 Measurements...
Page 240: ...240 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 15 Mask Testing...
Page 244: ...244 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter...
Page 256: ...256 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 17 Waveform Generator...
Page 274: ...274 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 19 Print Screens...
Page 308: ...308 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 21 Web Interface...
Page 326: ...326 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 22 Reference...