Horizontal Controls
To position the time reference (left, center, right)
Time reference is the reference point on the display for delay time (horizontal
position).
1
Press
[Horiz]
.
2
In the Horizontal Menu, press
Time Ref
; then, select:
•
Left
— the time reference is set to one major division from the left edge of the
display.
•
Center
— the time reference is set to the center of the display.
•
Right
— the time reference is set to one major division from the right edge of
the display.
A small hollow triangle (
∇
) at the top of the display grid marks the position of the
time reference. When delay time is set to zero, the trigger point indicator (
▼
)
overlays the time reference indicator.
The time reference position sets the initial position of the trigger event within
acquisition memory and on the display, with delay set to 0.
Turning the Horizontal scale (sweep speed) knob expands or contracts the
waveform about the time reference point (
∇
Turning the Horizontal position (
) knob in Normal mode (not Zoom) moves the
trigger point indicator (
▼
) to the left or right of the time reference point (
∇
). See
"To adjust the horizontal delay (position)"
Searching for Events
You can use the
[Search]
key and menu to search for Serial events on the analog
channels.
Setting up searches (see
on page 57) is similar to setting up
triggers.
Searches are different than triggers in that they use the measurement threshold
settings instead of trigger levels.
Summary of Contents for DSO-X 2014A
Page 1: ...Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide...
Page 8: ...8 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide...
Page 60: ...60 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Page 68: ...68 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 3 Vertical Controls...
Page 118: ...118 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 6 Digital Channels...
Page 124: ...124 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 7 Serial Decode...
Page 130: ...130 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 8 Display Settings...
Page 168: ...168 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 10 Triggers...
Page 176: ...176 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling...
Page 202: ...202 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 13 Cursors...
Page 226: ...226 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 14 Measurements...
Page 240: ...240 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 15 Mask Testing...
Page 244: ...244 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter...
Page 256: ...256 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 17 Waveform Generator...
Page 274: ...274 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 19 Print Screens...
Page 308: ...308 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 21 Web Interface...
Page 326: ...326 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 22 Reference...