Getting Started
1
Keysight InfiniiVision 2000 X-Series Oscilloscopes User's Guide
29
2
Connect the probe's retractable hook tip to the point of interest on the circuit or
device under test. Be sure to connect the probe ground lead to a ground point
on the circuit.
Input a Waveform
The first signal to input to the oscilloscope is the Demo 2, Probe Comp signal. This
signal is used for compensating probes.
1
Connect an oscilloscope probe from channel 1 to the
Demo 2
(Probe Comp)
terminal on the front panel.
2
Connect the probe's ground lead to the ground terminal (next to the
Demo 2
terminal).
CAUTION
Maximum input voltage at analog inputs
135 Vrms
CAUTION
When measuring voltages over 30 V, use a 10:1 probe.
CAUTION
Do not float the oscilloscope chassis
Defeating the ground connection and "floating" the oscilloscope chassis will probably
result in inaccurate measurements and may also cause equipment damage. The probe
ground lead is connected to the oscilloscope chassis and the ground wire in the power
cord. If you need to measure between two live points, use a differential probe with
sufficient dynamic range.
WARNING
Do not negate the protective action of the ground connection to the oscilloscope. The
oscilloscope must remain grounded through its power cord. Defeating the ground
creates an electric shock hazard.
Summary of Contents for DSO-X 2014A
Page 1: ...Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide...
Page 8: ...8 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide...
Page 60: ...60 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Page 68: ...68 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 3 Vertical Controls...
Page 118: ...118 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 6 Digital Channels...
Page 124: ...124 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 7 Serial Decode...
Page 130: ...130 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 8 Display Settings...
Page 168: ...168 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 10 Triggers...
Page 176: ...176 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling...
Page 202: ...202 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 13 Cursors...
Page 226: ...226 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 14 Measurements...
Page 240: ...240 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 15 Mask Testing...
Page 244: ...244 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter...
Page 256: ...256 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 17 Waveform Generator...
Page 274: ...274 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 19 Print Screens...
Page 308: ...308 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 21 Web Interface...
Page 326: ...326 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 22 Reference...