Serial Decode
7
Keysight InfiniiVision 2000 X-Series Oscilloscopes User's Guide
121
Before you can select a row or navigate through the Lister data, oscilloscope
acquisitions must be stopped.
4
Press the
[Single]
key (in the Run Control group on the front panel) to stop the
acquisition.
Pressing
[Single]
instead of
[Stop]
fills the maximum memory depth.
When zoomed out and viewing a large number of packets, the Lister may not
be able to display information for all packets. However, when you press the
[Single]
key the Lister will contain all on-screen serial decode information.
5
Press the
Scroll Lister
softkey and use the Entry knob to scroll through the data.
Time tags in the Time column indicate the event time relative to the trigger
point. The time tags of events that are shown in the waveform display area are
displayed with a dark background.
6
Press the
Zoom to Selection
softkey (or push the Entry knob) to center the
waveform display at the time associated with the selected Lister row and
automatically set the horizontal scale setting.
7
Press the
Undo Zoom
softkey to return to the horizontal scale and delay settings
before the last
Zoom to Selection
.
8
Press the
Options
softkey to open the Lister Options Menu. In this menu, you
can:
Summary of Contents for DSO-X 2014A
Page 1: ...Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide...
Page 8: ...8 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide...
Page 60: ...60 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Page 68: ...68 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 3 Vertical Controls...
Page 118: ...118 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 6 Digital Channels...
Page 124: ...124 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 7 Serial Decode...
Page 130: ...130 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 8 Display Settings...
Page 168: ...168 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 10 Triggers...
Page 176: ...176 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling...
Page 202: ...202 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 13 Cursors...
Page 226: ...226 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 14 Measurements...
Page 240: ...240 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 15 Mask Testing...
Page 244: ...244 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter...
Page 256: ...256 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 17 Waveform Generator...
Page 274: ...274 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 19 Print Screens...
Page 308: ...308 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 21 Web Interface...
Page 326: ...326 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 22 Reference...