Mask Testing
15
Keysight InfiniiVision 2000 X-Series Oscilloscopes User's Guide
233
To manually modify a mask file
You can manually modify a mask file that you created using the Automask
function.
1
Follow the steps 1-7 in
"To create a mask from a "golden" waveform
on page 227. Do not clear the mask after creating it.
2
Attach a USB mass storage device to the oscilloscope.
3
Press the
[Save/Recall]
key.
4
Press the
Save
softkey.
5
Press the
Format
softkey and select
Mask
.
6
Press the second softkey and select a destination folder on your USB mass
storage device.
7
Press the
Press to Save
softkey. This creates an ASCII text file that describes the
mask.
8
Remove the USB mass storage device and connect it to a PC.
9
Open the .msk file your created using a text editor (such as Wordpad).
10
Edit, save, and close the file.
The mask file contains the following sections:
• Mask File Identifier.
• Mask Title.
Reset Statistics
Note that statistics are also reset when:
•
Mask Test is switched on after being switched off.
•
Clear Mask softkey is pressed.
•
An Automask is created.
Additionally, the accumulated time counter is reset whenever the oscilloscope
is run after the acquisition was stopped.
Transparent
Enable Transparent mode to write measurement values and statistics on the
screen without a background. Disable the Transparent mode to show them with
a gray background. The Transparent setting affects mask test statistics,
measurement statistics, and reference waveform information display.
Clear Display
Clears acquisition data from the oscilloscope display.
Summary of Contents for DSO-X 2014A
Page 1: ...Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide...
Page 8: ...8 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide...
Page 60: ...60 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Page 68: ...68 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 3 Vertical Controls...
Page 118: ...118 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 6 Digital Channels...
Page 124: ...124 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 7 Serial Decode...
Page 130: ...130 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 8 Display Settings...
Page 168: ...168 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 10 Triggers...
Page 176: ...176 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling...
Page 202: ...202 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 13 Cursors...
Page 226: ...226 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 14 Measurements...
Page 240: ...240 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 15 Mask Testing...
Page 244: ...244 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter...
Page 256: ...256 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 17 Waveform Generator...
Page 274: ...274 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 19 Print Screens...
Page 308: ...308 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 21 Web Interface...
Page 326: ...326 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 22 Reference...