Trigger Mode/Coupling
To set the trigger holdoff
Trigger holdoff sets the amount of time the oscilloscope waits after a trigger
before re-arming the trigger circuitry.
Use the holdoff to trigger on repetitive waveforms that have multiple edges (or
other events) between waveform repetitions. You can also use holdoff to trigger on
the first edge of a burst when you know the minimum time between bursts.
For example, to get a stable trigger on the repetitive pulse burst shown below, set
the holdoff time to be >200 ns but <600 ns.
To set the trigger holdoff:
1
Press the
[Mode/Coupling]
key.
2
In the Trigger Mode and Coupling Menu, press the
Holdoff
softkey; then, turn
the Entry knob to increase or decrease the trigger holdoff time.
Trigger Holdoff
Operating Hints
The correct holdoff setting is typically slightly less than one repetition of the
waveform. Set the holdoff to this time to generate a unique trigger point for a
repetitive waveform.
Changing the time base settings does not affect the trigger holdoff time.
With Keysight's MegaZoom technology, you can press
[Stop]
, then pan and zoom
through the data to find where the waveform repeats. Measure this time using
cursors; then, set the holdoff.
External Trigger Input
The external trigger input can be used as a source in several of the trigger types.
The external trigger BNC input is on the rear panel and is labeled
EXT TRIG IN
.
200 ns
Oscilloscope triggers here
Holdoff
600 ns
Summary of Contents for DSO-X 2014A
Page 1: ...Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide...
Page 8: ...8 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide...
Page 60: ...60 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Page 68: ...68 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 3 Vertical Controls...
Page 118: ...118 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 6 Digital Channels...
Page 124: ...124 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 7 Serial Decode...
Page 130: ...130 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 8 Display Settings...
Page 168: ...168 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 10 Triggers...
Page 176: ...176 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling...
Page 202: ...202 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 13 Cursors...
Page 226: ...226 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 14 Measurements...
Page 240: ...240 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 15 Mask Testing...
Page 244: ...244 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter...
Page 256: ...256 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 17 Waveform Generator...
Page 274: ...274 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 19 Print Screens...
Page 308: ...308 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 21 Web Interface...
Page 326: ...326 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 22 Reference...