I2C/SPI Triggering and Serial Decode
24
Keysight InfiniiVision 2000 X-Series Oscilloscopes User's Guide
351
2
In the Search Menu, press the
Search
softkey; then, turn the Entry knob to select
the serial slot (Serial 1 or Serial 2) on which the I2C signal is being decoded.
3
Press
Search
; then, select from these options:
•
Missing Acknowledge
— finds SDA data is high during any Ack SCL clock bit.
•
Address with no Ack
— finds when the acknowledge of the selected address
field is false. The R/W bit is ignored.
•
Restart
— finds when another start condition occurs before a stop condition.
•
EEPROM Data Read
— finds EEPROM control byte value 1010xxx on the SDA
line, followed by a Read bit and an Ack bit. It then looks for the data value
and qualifier set by the Data is softkey and the Data softkeys.
•
Frame(Start:Address7:Read:Ack:Data)
— finds a read frame on the 17th clock
edge if all bits in the pattern match.
•
Frame(Start:Address7:Write:Ack:Data)
— finds a write frame on the 17th clock
edge if all bits in the pattern match.
•
Frame(Start:Address7:Read:Ack:Data:Ack:Data2)
— finds a read frame on the 26th
clock edge if all bits in the pattern match.
•
Frame(Start:Address7:Write:Ack:Data:Ack:Data2)
— finds a write frame on the
26th clock edge if all bits in the pattern match.
For more information on searching data, see
For more information on using the
[Navigate]
key and controls, see
Setup for SPI Signals
Serial Peripheral Interface (SPI) signals setup consists of connecting the
oscilloscope to a clock, MOSI data, MISO data, and framing signal, then setting
the threshold voltage level for each input channel, and finally specifying any other
signal parameters.
To set up the oscilloscope to capture SPI signals, use the
Signals
softkey which
appears in the Serial Decode Menu:
1
Press
[Label]
to turn on labels.
2
Press
[Serial]
.
3
Press the
Mode
softkey; then, select
SPI
trigger type.
Summary of Contents for DSO-X 2014A
Page 1: ...Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide...
Page 8: ...8 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide...
Page 60: ...60 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Page 68: ...68 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 3 Vertical Controls...
Page 118: ...118 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 6 Digital Channels...
Page 124: ...124 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 7 Serial Decode...
Page 130: ...130 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 8 Display Settings...
Page 168: ...168 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 10 Triggers...
Page 176: ...176 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling...
Page 202: ...202 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 13 Cursors...
Page 226: ...226 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 14 Measurements...
Page 240: ...240 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 15 Mask Testing...
Page 244: ...244 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter...
Page 256: ...256 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 17 Waveform Generator...
Page 274: ...274 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 19 Print Screens...
Page 308: ...308 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 21 Web Interface...
Page 326: ...326 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 22 Reference...