Digital Voltmeter
The DVM makes accurate RMS measurements when the signal frequency is
between 20 Hz and 100 kHz. When the signal frequency is outside this range,
"<BW Limit?" or ">BW Limit?" appears in the DVM display to caution you about
inaccurate RMS measurement results.
To use the digital voltmeter:
1
Press the
[Analyze]
key.
2
Press
Features
; then, select
Digital Voltmeter
.
3
Press
Features
again to enable the DVM measurements.
4
Press the
Source
softkey and turn the Entry knob to select the analog channel
on which digital voltmeter (DVM) measurements are made.
The selected channel does not have to be on (displaying a waveform) in order
for DVM measurements to be made.
5
Press the
Mode
softkey and turn the Entry knob to select the digital voltmeter
(DVM) mode:
•
AC RMS
— displays the root-mean-square value of the acquired data, with
the DC component removed.
•
DC
— displays the DC value of the acquired data.
•
DC RMS
— displays the root-mean-square value of the acquired data.
Summary of Contents for DSO-X 2014A
Page 1: ...Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide...
Page 8: ...8 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide...
Page 60: ...60 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Page 68: ...68 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 3 Vertical Controls...
Page 118: ...118 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 6 Digital Channels...
Page 124: ...124 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 7 Serial Decode...
Page 130: ...130 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 8 Display Settings...
Page 168: ...168 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 10 Triggers...
Page 176: ...176 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling...
Page 202: ...202 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 13 Cursors...
Page 226: ...226 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 14 Measurements...
Page 240: ...240 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 15 Mask Testing...
Page 244: ...244 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter...
Page 256: ...256 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 17 Waveform Generator...
Page 274: ...274 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 19 Print Screens...
Page 308: ...308 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 21 Web Interface...
Page 326: ...326 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 22 Reference...