Math Waveforms
4
Keysight InfiniiVision 2000 X-Series Oscilloscopes User's Guide
71
To perform transforms or filters on an arithmetic operation
To perform a transform function (see
on page 74) or filter (see
on page 90) on the add, subtract, or multiply arithmetic operations:
1
Press the
Function
softkey and select
g(t): Internal
.
2
Use the
Operator
,
Source 1
, and
Source 2
softkeys to set up an arithmetic
operation.
3
Press the
Function
softkey and select
f(t): Displayed
.
4
Use the
Operator
softkey to select a transform function or filter.
5
Press the
Source 1
softkey and select
g(t)
as the source. Note that
g(t)
is only
available when you select a transform function in the previous step.
To adjust the math waveform scale and offset
1
Make sure the multiplexed scale and position knobs to the right of the
[Math]
key are selected for the math waveform.
If the arrow to the left of the
[Math]
key is not illuminated, press the key.
2
Use the multiplexed scale and position knobs just to the right of the
[Math]
key
to re-size and re-position the math waveform.
See Also
•
Units for Math Waveforms
Units for each input channel can be set to Volts or Amps using the
Units
softkey in
the channel's Probe Menu. Units for math function waveforms are:
NOTE
Math Scale and Offset are Set Automatically
Any time the currently displayed math function definition is changed, the function is
automatically scaled for optimum vertical scale and offset. If you manually set scale and offset
for a function, select a new function, then select the original function, the original function will
be automatically rescaled.
Summary of Contents for DSO-X 2014A
Page 1: ...Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide...
Page 8: ...8 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide...
Page 60: ...60 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Page 68: ...68 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 3 Vertical Controls...
Page 118: ...118 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 6 Digital Channels...
Page 124: ...124 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 7 Serial Decode...
Page 130: ...130 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 8 Display Settings...
Page 168: ...168 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 10 Triggers...
Page 176: ...176 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling...
Page 202: ...202 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 13 Cursors...
Page 226: ...226 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 14 Measurements...
Page 240: ...240 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 15 Mask Testing...
Page 244: ...244 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter...
Page 256: ...256 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 17 Waveform Generator...
Page 274: ...274 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 19 Print Screens...
Page 308: ...308 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 21 Web Interface...
Page 326: ...326 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 22 Reference...