Digital Channels
6
Keysight InfiniiVision 2000 X-Series Oscilloscopes User's Guide
115
The impedance plots for the two models are shown in these figures. By comparing
the two plots, you can see that both the series tip resistor and the cable's
characteristic impedance extend the input impedance significantly. The stray tip
capacitance, which is generally small (1 pF), sets the final break point on the
impedance chart.
The logic probes are represented by the high-frequency circuit model shown
above. They are designed to provide as much series tip resistance as possible.
Stray tip capacitance to ground is minimized by the proper mechanical design of
the probe tip assembly. This provides the maximum input impedance at high
frequencies.
Probe Grounding
A probe ground is the low-impedance path for current to return to the source from
the probe. Increased length in this path will, at high frequencies, create large
common mode voltages at the probe input. The voltage generated behaves as if
this path were an inductor according to the equation:
Figure 23
Impedance versus Frequency for Both Probe Circuit Models
100 k
10 k
1 k
100
10
1
10 kHz
100 kHz
1 MHz
10 MHz
100 MHz
1 GHz
High
Frequency
Model
Typical
Model
Frequency
Impedance
Summary of Contents for DSO-X 2014A
Page 1: ...Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide...
Page 8: ...8 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide...
Page 60: ...60 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Page 68: ...68 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 3 Vertical Controls...
Page 118: ...118 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 6 Digital Channels...
Page 124: ...124 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 7 Serial Decode...
Page 130: ...130 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 8 Display Settings...
Page 168: ...168 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 10 Triggers...
Page 176: ...176 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling...
Page 202: ...202 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 13 Cursors...
Page 226: ...226 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 14 Measurements...
Page 240: ...240 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 15 Mask Testing...
Page 244: ...244 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter...
Page 256: ...256 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 17 Waveform Generator...
Page 274: ...274 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 19 Print Screens...
Page 308: ...308 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 21 Web Interface...
Page 326: ...326 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 22 Reference...