Triggers
10
Keysight InfiniiVision 2000 X-Series Oscilloscopes User's Guide
147
•
0
sets the pattern to zero (low) on the selected channel. A low is a voltage
level that is less than the channel's trigger level or threshold level.
•
1
sets the pattern to 1 (high) on the selected channel. A high is a voltage
level that is greater than the channel's trigger level or threshold level.
•
X
sets the pattern to don't care on the selected channel. Any channel set to
don't care is ignored and is not used as part of the pattern. However, if all
channels in the pattern are set to don't care, the oscilloscope will not trigger.
•
The rising edge ( ) or falling edge ( ) softkey sets the pattern to an edge
on the selected channel. Only one rising or falling edge can be specified in
the pattern. When an edge is specified, the oscilloscope will trigger at the
edge specified if the pattern set for the other channels is true.
If no edge is specified, the oscilloscope will trigger on the last edge that
makes the pattern true.
NOTE
Specifying an Edge in a Pattern
You are allowed to specify only one rising or falling edge term in the pattern. If you define an
edge term, then select a different channel in the pattern and define another edge term, the
previous edge definition is changed to a don't care.
Summary of Contents for DSO-X 2014A
Page 1: ...Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide...
Page 8: ...8 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide...
Page 60: ...60 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Page 68: ...68 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 3 Vertical Controls...
Page 118: ...118 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 6 Digital Channels...
Page 124: ...124 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 7 Serial Decode...
Page 130: ...130 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 8 Display Settings...
Page 168: ...168 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 10 Triggers...
Page 176: ...176 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling...
Page 202: ...202 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 13 Cursors...
Page 226: ...226 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 14 Measurements...
Page 240: ...240 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 15 Mask Testing...
Page 244: ...244 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter...
Page 256: ...256 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 17 Waveform Generator...
Page 274: ...274 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 19 Print Screens...
Page 308: ...308 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 21 Web Interface...
Page 326: ...326 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 22 Reference...