12
Acquisition Control
Oscilloscope Bandwidth Required
The oscilloscope bandwidth required to accurately measure a signal is primarily
determined by the signal's rise time, not the signal's frequency. You can use these
steps to calculate the oscilloscope bandwidth required:
1
Determine the fastest edge speeds.
You can usually obtain rise time information from published specifications for
devices used in your designs.
2
Compute the maximum "practical" frequency component.
From Dr. Howard W. Johnson's book,
High-Speed Digital Design – A
Handbook of Black Magic
, all fast edges have an infinite spectrum of frequency
components. However, there is an inflection (or "knee") in the frequency
spectrum of fast edges where frequency components higher than f
knee
are
insignificant in determining the shape of the signal.
f
knee
= 0.5 / signal rise time (based on 10% - 90% thresholds)
f
knee
= 0.4 / signal rise time (based on 20% - 80% thresholds)
3
Use a multiplication factor for the required accuracy to determine the
oscilloscope bandwidth required.
See Also
Choosing an Oscilloscope with the Right Bandwidth for your Application
, Keysight
Application Note 1588
(
http://literature.cdn.keysight.com/litweb/pdf/5989-5733EN.pdf
)
Memory Depth and Sample Rate
The number of points of oscilloscope memory is fixed, and there is a maximum
sample rate associated with oscilloscope's analog-to-digital converter; however,
the actual sample rate is determined by the time of the acquisition (which is set
according to the oscilloscope's horizontal time/div scale).
Required accuracy
Oscilloscope bandwidth required
20%
f
BW
= 1.0 x f
knee
10%
f
BW
= 1.3 x f
knee
3%
f
BW
= 1.9 x f
knee
Summary of Contents for DSO-X 2014A
Page 1: ...Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide...
Page 8: ...8 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide...
Page 60: ...60 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Page 68: ...68 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 3 Vertical Controls...
Page 118: ...118 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 6 Digital Channels...
Page 124: ...124 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 7 Serial Decode...
Page 130: ...130 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 8 Display Settings...
Page 168: ...168 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 10 Triggers...
Page 176: ...176 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling...
Page 202: ...202 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 13 Cursors...
Page 226: ...226 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 14 Measurements...
Page 240: ...240 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 15 Mask Testing...
Page 244: ...244 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter...
Page 256: ...256 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 17 Waveform Generator...
Page 274: ...274 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 19 Print Screens...
Page 308: ...308 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 21 Web Interface...
Page 326: ...326 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 22 Reference...