184
Keysight InfiniiVision 2000 X-Series Oscilloscopes User's Guide
12
Acquisition Control
•
Averaging
— at all time/div settings, the specified number of triggers are
averaged together. Use this mode for reducing noise and increasing
resolution of periodic signals without bandwidth or rise time degradation.
•
High Resolution
— at slower time/div settings, all samples in the effective
sample period are averaged and the average value is stored. Use this mode
for reducing random noise. See
"High Resolution Acquisition Mode"
Normal Acquisition Mode
In Normal mode at slower time/div settings, extra samples are decimated (in other
words, some are thrown away). This mode yields the best display for most
waveforms.
Peak Detect Acquisition Mode
In Peak Detect mode, at slower time/div settings when decimation would normally
occur, minimum and maximum value samples are kept in order to capture
infrequent and narrow events (at the expense of exaggerating any noise). This
mode displays all pulses that are at least as wide as the sample period.
For InfiniiVision 2000 X-Series oscilloscopes, which have a maximum sample rate
of 2 GSa/s, a sample is taken every 500 ps (sample period).
See Also
•
"Glitch or Narrow Pulse Capture"
•
"Using Peak Detect Mode to Find a Glitch"
Glitch or Narrow Pulse Capture
A glitch is a rapid change in the waveform that is usually narrow as compared to
the waveform. Peak detect mode can be used to more easily view glitches or
narrow pulses. In peak detect mode, narrow glitches and sharp edges are
displayed more brightly than when in Normal acquire mode, making them easier
to see.
To characterize the glitch, use the cursors or the automatic measurement
capabilities of the oscilloscope.
Summary of Contents for DSO-X 2014A
Page 1: ...Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide...
Page 8: ...8 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide...
Page 60: ...60 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Page 68: ...68 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 3 Vertical Controls...
Page 118: ...118 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 6 Digital Channels...
Page 124: ...124 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 7 Serial Decode...
Page 130: ...130 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 8 Display Settings...
Page 168: ...168 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 10 Triggers...
Page 176: ...176 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling...
Page 202: ...202 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 13 Cursors...
Page 226: ...226 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 14 Measurements...
Page 240: ...240 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 15 Mask Testing...
Page 244: ...244 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter...
Page 256: ...256 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 17 Waveform Generator...
Page 274: ...274 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 19 Print Screens...
Page 308: ...308 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 21 Web Interface...
Page 326: ...326 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 22 Reference...