Triggers
10
Keysight InfiniiVision 2000 X-Series Oscilloscopes User's Guide
145
When triggering on a positive pulse, the trigger will occur on the high to low
transition of the pulse if the qualifying condition is true. When triggering on a
negative pulse, the trigger will occur on the low to high transition of the pulse if
the qualifying condition is true.
6
Press the qualifier softkey (
< > ><
) to select the time qualifier.
The Qualifier softkey can set the oscilloscope to trigger on a pulse width that is:
•
Less than a time value (
<
).
For example, for a positive pulse, if you set t<10 ns:
•
Greater than a time value (
>
).
For example, for a positive pulse, if you set t>10 ns:
•
Within a range of time values (
><
).
For example, for a positive pulse, if you set t>10 ns and t<15 ns:
7
Select the qualifier time set softkey (
<
or
>
), then rotate the Entry knob to set
the pulse width qualifier time.
The qualifiers can be set as follows:
•
2 ns to 10 s for > or < qualifier (5 ns to 10 s for 350 MHz bandwidth models).
•
10 ns to 10 s for >< qualifier, with minimum difference of 5 ns between upper
and lower settings.
Pulse width
trigger < qualifier
time set softkey
• When the less than (
<
) qualifier is selected, the Entry knob sets the oscilloscope
to trigger on a pulse width less than the time value displayed on the softkey.
10 ns
10 ns
Trigger
10 ns
10 ns
Trigger
10 ns
15 ns
Trigger
12 ns
Summary of Contents for DSO-X 2014A
Page 1: ...Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide...
Page 8: ...8 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide...
Page 60: ...60 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Page 68: ...68 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 3 Vertical Controls...
Page 118: ...118 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 6 Digital Channels...
Page 124: ...124 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 7 Serial Decode...
Page 130: ...130 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 8 Display Settings...
Page 168: ...168 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 10 Triggers...
Page 176: ...176 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling...
Page 202: ...202 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 13 Cursors...
Page 226: ...226 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 14 Measurements...
Page 240: ...240 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 15 Mask Testing...
Page 244: ...244 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter...
Page 256: ...256 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 17 Waveform Generator...
Page 274: ...274 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 19 Print Screens...
Page 308: ...308 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 21 Web Interface...
Page 326: ...326 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 22 Reference...