140
Keysight InfiniiVision 2000 X-Series Oscilloscopes User's Guide
10
Triggers
Edge Trigger
The Edge trigger type identifies a trigger by looking for a specified edge (slope)
and voltage level on a waveform. You can define the trigger source and slope in
this menu. The trigger type, source, and level (if applicable) are displayed in the
upper-right corner of the display.
1
On the front panel, in the Trigger section, press the
[Trigger]
key.
2
In the Trigger Menu, press the
Trigger
softkey, and use the Entry knob to select
Edge
.
3
Select the trigger source:
•
Analog channel,
1
to the number of channels
•
Digital channel (on mixed-signal oscilloscopes),
D0
to the number of digital
channels minus one.
•
External
— triggers on the rear panel EXT TRIG IN signal.
•
Line
— triggers at the 50% level of the rising or falling edge of the AC power
source signal.
•
WaveGen
— triggers at the 50% level of the rising edge of the waveform
generator output signal. (Not available when the DC or Noise waveforms are
selected.)
You can choose a channel that is turned off (not displayed) as the source for the
edge trigger.
The selected trigger source is indicated in the upper-right corner of the display
next to the slope symbol:
•
1
through
4
= analog channels.
•
D0
through
Dn
= digital channels.
•
E
= External trigger input.
•
L
= Line trigger.
•
W
= Waveform generator.
4
Press the
Slope
softkey and select rising edge, falling edge, alternating edges,
or either edge (depending on the selected source). The selected slope is
displayed in the upper-right corner of the display.
Summary of Contents for DSO-X 2014A
Page 1: ...Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide...
Page 8: ...8 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide...
Page 60: ...60 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Page 68: ...68 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 3 Vertical Controls...
Page 118: ...118 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 6 Digital Channels...
Page 124: ...124 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 7 Serial Decode...
Page 130: ...130 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 8 Display Settings...
Page 168: ...168 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 10 Triggers...
Page 176: ...176 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling...
Page 202: ...202 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 13 Cursors...
Page 226: ...226 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 14 Measurements...
Page 240: ...240 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 15 Mask Testing...
Page 244: ...244 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter...
Page 256: ...256 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 17 Waveform Generator...
Page 274: ...274 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 19 Print Screens...
Page 308: ...308 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 21 Web Interface...
Page 326: ...326 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 22 Reference...