15 Mask Testing
To create a mask from a "golden" waveform (Automask) / 227
Mask Test Setup Options / 229
Mask Statistics / 232
To manually modify a mask file / 233
Building a Mask File / 236
One way to verify a waveform's compliance to a particular set of parameters is to
use mask testing. A mask defines a region of the oscilloscope's display in which
the waveform must remain in order to comply with chosen parameters.
Compliance to the mask is verified point-by-point across the display. Mask test
operates on displayed analog channels; it does not operate on channels that are
not displayed.
Mask test is a license-enabled feature.
To create a mask from a "golden" waveform (Automask)
A golden waveform meets all chosen parameters, and it is the waveform to which
all others will be compared.
1
Configure the oscilloscope to display the golden waveform.
2
Press the
[Analyze]
key.
3
Press
Features
; then, select
Mask Test
.
4
Press
Features
again to enable mask testing.
Summary of Contents for DSO-X 2014A
Page 1: ...Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide...
Page 8: ...8 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide...
Page 60: ...60 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Page 68: ...68 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 3 Vertical Controls...
Page 118: ...118 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 6 Digital Channels...
Page 124: ...124 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 7 Serial Decode...
Page 130: ...130 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 8 Display Settings...
Page 168: ...168 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 10 Triggers...
Page 176: ...176 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling...
Page 202: ...202 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 13 Cursors...
Page 226: ...226 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 14 Measurements...
Page 240: ...240 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 15 Mask Testing...
Page 244: ...244 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter...
Page 256: ...256 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 17 Waveform Generator...
Page 274: ...274 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 19 Print Screens...
Page 308: ...308 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 21 Web Interface...
Page 326: ...326 Keysight InfiniiVision 2000 X Series Oscilloscopes User s Guide 22 Reference...