M-2000® Hardware Manual
©2023 J.A. Woollam Co.
59
Figure 4-10. Nomenclature.
Source Unit Alignment
The lamp is housed inside the Source Unit, which is mounted to the system with
tip-tilt adjustment (Source Tilt Stage) to manipulate the beam direction. The
source beam alignment is verified in the straight-through position to ensure that
it is aligned on the Receiver Unit aperture.
1.
Select
Hardware
>
Controls
>
System
>‘Angle’. Choose an Angle of 90°. The
base will move to the straight-through position.
Figure 4-11. Angle button.
Source Unit
Source
Tilt stage
Receiver
Tilt stage
Receiver Unit
Sample
stage
Vacuum
switch