128
©2023 J.A. Woollam Co.
M-2000® Hardware Manual
b.
Z Align/Imaging Camera.
If the camera image is not in focus,
loosen the lock-down on each side and adjust the Z position of the
camera by turning the adjustment screw. Tighten the lock-downs.
Figure 8-6. Z Align/Imaging Camera Z position adjustment.
i.
If necessary,
translate
the Z/Imaging Camera to center the
beam on the screen. Remount the business card and
perform Intensity-based Z alignment again or flip
calibration wafer over to rough side.
Figure 8-7. Z Align/Imaging Camera translation adjustment.
Z position (focus)
Loosen lock-downs and Translate