M-2000® Hardware Manual
Figure 10-10. Reflective Mirror replacement.
4.
Align the mirror using the same instructions from the lamp replacement
section above. This can be done at the same time the lamp is changed.
10.3
QTH Lamp (M-2000V & VI)
in FQTH-100
This section describes installation/replacement, alignment, and cleaning of the
lamp used in the FQTH-100 Light Source (for M-2000V and VI systems).
The QTH lamp must be replaced after a certain period of time. As a lamp
approaches the end of its useful life, power levels can fluctuate and spectral
intensity can be reduced, both result in bad or inaccurate data.
LAMP
MODEL #
COLOR
TEMP.
RATING
TYPICAL
LIFETIME
ILT
L7417
3000°K
2000 hours
3000 hours
Oriel
6332
3300°K
50 hours
300 hours
Table 10-4. QTH lamps and typical lifetime.
Lamp Installation/Replacement
1.
Turn off the M-2000 Detector and Lamp Power Supply Module. Allow
lamp(s) to cool for 2 hours.
2.
Put on latex gloves and safety goggles.
3.
Remove (2) SHC screws (2.5mm Allen wrench) to remove the cover.
Mirror Cover Plate Screws
Retainer Clip