©2023 J.A. Woollam Co.
M-2000® Hardware Manual
Figure 4-43. Alignment Screen – detector selection.
5.
The cross hair will not likely be aligned. We will adjust the sample stage
to align the cross hair, effectively matching the Sample Stage to the
System Alignment.
6.
Align the sample Tip/Tilt by pressing the Up/Down/Left/Right Arrow keys
on keyboard. Tip and tilt the sample stage until the cross hair is centered
on the screen. You will have to adjust the Tilt Stage Motor Increment to
reduce the steps to get “fine” adjustment of the motors.
Figure 4-44. Adjusting tilt stage motor increment.
a.
If the Receiver Unit has the Alignment Camera, no gray intensity
will appear. If the beam is not entering the Receiver Unit at all, it
will display ‘Out of Range.’
Figure 4-45. Cross hair example – Receiver Camera installed.