M-2000® Hardware Manual
©2023 J.A. Woollam Co.
75
2.
Select
Hardware
>
Controls
>
System
>‘Z-Stage’ and move to +0.55mm as a
starting position for the thickness of the Calibration Wafer.
3.
Select
Hardware
>
Controls
>
System
>‘Angle’ and move to 65°.
Figure 4-40.
Hardware
>
Controls
>
System.
4.
Select
Hardware
>
Controls
>
General
>‘Align Sample’. The software will
ask “Would you like to perform a full sample alignment?” Select ‘No’.
There are two alignment detectors on this system. In the Alignment
Screen, change the Detector to "Receiver".
Figure 4-41. Align Sample button.
Figure 4-42. Sample Alignment dialog box.
Note:
The following alignment screen is a generic screen shot. Refer to
the system test report for your specific intensity values.