114
©2023 J.A. Woollam Co.
M-2000® Hardware Manual
Angle Offset Correction
Due to alignment sensitivity, the angle offset may need to be corrected each
time the Focusing in installed. See section 11.6 Angle of Incidence Offset
Correction under the Performance Checks chapter. If equipped with Camera
Alignment, this will be done automatically when the Focusing is installed.
6.6
Removal
To remove the Focusing Optics, select
Hardware
>
Controls
>
System
>‘Remove
Focus Probes.’ The base will move to the maximum Z height position,
maximizing the space between the Focusing Optics and sample stage, and
minimizing risk of hitting the sample stage during removal and damaging the
optics.
A message will appear letting you know it is safe to remove the Focusing Probes.
Loosen the three captive thumb screws for each probe to remove. Store the
Focusing Probes with protective caps over the lenses.
Figure 6-25. Remove Focus Probes button.
Figure 6-26. Focus Probe Removal message box.