standard SiO2 on Si reference sample supplied with the instrument
should be used for this procedure.
Note:
A System Check should be performed any time the performance of
the instrument is in question.
1.
Select
Hardware
>
Controls
>
General
>‘System Check’
2.
Click ‘Yes’ to start the System Check.
Figure 4-62. System Check message box.
3.
When prompted, mount the calibration wafer (~25nm SiO2 on Si,
supplied with the instrument).
Figure 4-63. Mount Calibration Wafer.
4.
CompleteEASE will automatically open the sample alignment screen.
Align the sample as described in the Sample Alignment sections above.
Figure 4-64. Alignment screen.