M-2000® Hardware Manual
©2023 J.A. Woollam Co.
83
4.
Select
Hardware
>
Controls
>
General
>‘Align Sample’. When prompted to
perform a full system alignment, select ‘Yes’.
Figure 4-57. Sample Alignment message box.
5.
Under the Sample Tilt Alignment drop down menu, select Manual. Under
the Sample Height Alignment drop down menu, select Automatic-Quick.
Enter the sample thickness into the box. (Cal. Wafer ≈ 0.55mm). The
Alignment Angle should be 65°. Press ‘Ok’.
Figure 4-58. Sample Alignment dialog box.
6.
Adjust the sample tilt stage knobs until the cross hair is centered.
Note:
The following alignment screen is a generic screen shot. Refer to
the system test report for your specific intensity values.
Figure 4-59. Cross hair centered.
a.
If the Receiver Unit has the Alignment Camera, no gray intensity
will appear. If the beam is not entering the Receiver Unit at all, it
will display ‘Out of Range.’