I
Making Measurements
Measuring AM Delay (Option
or
This example is an AM delay measurement of a frequency converter.
view the measurement trace, press
.
2. lb interpret the AM delay measurement, refer to Figure 3-18.
a. Note that the vertical axis is displaying time rather than power as
in previous example measurements. The AM delay measurement
measures the time required for power to travel through the DUT at
various frequencies. The measurement trace will be noisier as the
power level is attenuated by the DUT.
b. Since delay is proportional to the derivative of phase, flat (constant)
delay indicates linear phase. Delay measurements are typically
performed to measure the deviation from linear phase. Deviation from
linear phase (flat delay) would indicate that the DUT is distorting the
signal.
3-46
Summary of Contents for 8712C
Page 1: ...I I User s Guide HP 8712C and HP 8714C RF Network Analyzers I ...
Page 17: ...I 1 Installing the Analyzer ...
Page 35: ...I 2 Getting Started ...
Page 54: ...I 3 Making Measurements ...
Page 81: ......
Page 119: ...4 Using Instrument Functions ...
Page 182: ...Using Instrument Functions Customizing the Display Figure 4 24 Expanded Display 4 64 ...
Page 223: ...5 Optimizing Measurements ...
Page 244: ...6 Calibrating for Increased Measurement Accuracy ...
Page 280: ...7 Automating Measurements ...
Page 298: ......
Page 362: ...8 Front Rear Panel ...
Page 373: ......
Page 381: ......
Page 386: ... 9 w I Suftkey Reference ...
Page 477: ...I 10 Specifications and Characteristics ...
Page 501: ...11 Safety and Regulatory Information ...
Page 508: ...I 12 preset State and Memory Allocation ...
Page 527: ...Index ...
Page 559: ...Y X 9 91 YYW m m IiH rn 9 91 Z zeroing detectors 6 15 Index 33 ...