Check the Calibration
Your analyzer has a calibration check feature that can compute and display
corrected measurement uncertainties (residual errors) that apply to the
current instrument settings and calibrations.
During a calibration check, you are prompted to connect calibration standards
to your measurement ports. These standards are measured with the current
calibration corrections applied and the residual errors are then calculated and
can be displayed.
The calibration check feature can be a useful tool to help you troubleshoot
poor measurements and to help you determine how often to perform
calibrations on your analyzer.
Troubleshooting with
If you suspect that you may not be making valid measurements, you can use
Calibration Check
the calibration check to confirm that your current measurement calibration is
valid or. to reveal that it is faulty. Using a different set of calibration standards
for the calibration check than the ones you used for the initial calibration will
help you to rule out the possibility of degraded or faulty cal standards.
Using Calibration Check
Using the calibration check feature can help you determine the optimum
to Determine
calibration interval for your analyzer in your particular environment.
Calibration Interval
Calibrations can degrade over time due to temperature drift, connector wear,
cable movements, etc.
use the calibration check feature to determine the best calibration interval
for your measurements:
1. Perform the type of calibration desired.
2. Determine the residual errors by doing a calibration check.
6-26
Summary of Contents for 8712C
Page 1: ...I I User s Guide HP 8712C and HP 8714C RF Network Analyzers I ...
Page 17: ...I 1 Installing the Analyzer ...
Page 35: ...I 2 Getting Started ...
Page 54: ...I 3 Making Measurements ...
Page 81: ......
Page 119: ...4 Using Instrument Functions ...
Page 182: ...Using Instrument Functions Customizing the Display Figure 4 24 Expanded Display 4 64 ...
Page 223: ...5 Optimizing Measurements ...
Page 244: ...6 Calibrating for Increased Measurement Accuracy ...
Page 280: ...7 Automating Measurements ...
Page 298: ......
Page 362: ...8 Front Rear Panel ...
Page 373: ......
Page 381: ......
Page 386: ... 9 w I Suftkey Reference ...
Page 477: ...I 10 Specifications and Characteristics ...
Page 501: ...11 Safety and Regulatory Information ...
Page 508: ...I 12 preset State and Memory Allocation ...
Page 527: ...Index ...
Page 559: ...Y X 9 91 YYW m m IiH rn 9 91 Z zeroing detectors 6 15 Index 33 ...