I
Making Measurements
Measuring Group Delay
1.
2.
3.
Since we are measuring the transmission group delay, a transmission
calibration can be performed to improve accuracy
The instrument prompts you to connect four standards-open, short, load,
and through cable-as shown below.
NETWORK ANALYZER
N E T W O R K A N A L Y Z E R
R F O U T
R F O U T
R F I N
OPEN
SHORT
LOAD
Open, Short, load Connections
Through Cable Connection
4. Press Measure
after connecting each standard.
5. The analyzer will measure each standard and then calculate the new
calibration coefficients. The message “Calibration complete. will
appear for a few seconds when the analyzer is done calculating the new
error correction array.
6. The calibration may be saved in memory or on a disk for later use if you
wish. See Chapter 6 for information on saving calibrations.
3-51
Summary of Contents for 8712C
Page 1: ...I I User s Guide HP 8712C and HP 8714C RF Network Analyzers I ...
Page 17: ...I 1 Installing the Analyzer ...
Page 35: ...I 2 Getting Started ...
Page 54: ...I 3 Making Measurements ...
Page 81: ......
Page 119: ...4 Using Instrument Functions ...
Page 182: ...Using Instrument Functions Customizing the Display Figure 4 24 Expanded Display 4 64 ...
Page 223: ...5 Optimizing Measurements ...
Page 244: ...6 Calibrating for Increased Measurement Accuracy ...
Page 280: ...7 Automating Measurements ...
Page 298: ......
Page 362: ...8 Front Rear Panel ...
Page 373: ......
Page 381: ......
Page 386: ... 9 w I Suftkey Reference ...
Page 477: ...I 10 Specifications and Characteristics ...
Page 501: ...11 Safety and Regulatory Information ...
Page 508: ...I 12 preset State and Memory Allocation ...
Page 527: ...Index ...
Page 559: ...Y X 9 91 YYW m m IiH rn 9 91 Z zeroing detectors 6 15 Index 33 ...