Using Instrument Functions
Using Markers
In marker delta mode, a reference marker is placed at the active marker
position. All marker values are then displayed in reference to this delta
marker. When the amplitude of the measurement trace changes, the
reference marker
also changes. The delta marker is represented by
a triangle pointing up (A) with a delta symbol (A) below it rather than a
number.
Limit testing can be performed on delta amplitude and delta frequency using
specific markers. See
Use Marker Limit Functions, later in this chapter.
follow along with this example, set up the instrument for a transmission
measurement of the
that was shipped with your instrument.
Press
Search
to
move marker 1 (the active marker) to the maximum value on the trace.
Press
Prior
ON
to place a reference marker at the active marker position.
Press Prior
to
place marker 1 to the minimum point on the trace.
4. The difference between the markers’ frequency and amplitude value is
shown in the upper-right corner of the analyzer screen.
Press
Menu Prior
2: . Use the front panel knob to move
marker 2 towards the right-hand side of the display screen. Note that the
marker 2 values are also in reference to the delta marker.
4-27
Summary of Contents for 8712C
Page 1: ...I I User s Guide HP 8712C and HP 8714C RF Network Analyzers I ...
Page 17: ...I 1 Installing the Analyzer ...
Page 35: ...I 2 Getting Started ...
Page 54: ...I 3 Making Measurements ...
Page 81: ......
Page 119: ...4 Using Instrument Functions ...
Page 182: ...Using Instrument Functions Customizing the Display Figure 4 24 Expanded Display 4 64 ...
Page 223: ...5 Optimizing Measurements ...
Page 244: ...6 Calibrating for Increased Measurement Accuracy ...
Page 280: ...7 Automating Measurements ...
Page 298: ......
Page 362: ...8 Front Rear Panel ...
Page 373: ......
Page 381: ......
Page 386: ... 9 w I Suftkey Reference ...
Page 477: ...I 10 Specifications and Characteristics ...
Page 501: ...11 Safety and Regulatory Information ...
Page 508: ...I 12 preset State and Memory Allocation ...
Page 527: ...Index ...
Page 559: ...Y X 9 91 YYW m m IiH rn 9 91 Z zeroing detectors 6 15 Index 33 ...