I
Using Instrument Functions
Using Markers
Using the Next Peak and Next Min Functions.
As explained previously, pressing
and
will
place a marker on the maximum and minimum points on the measurement
trace, respectively. You can search for the next highest or lowest point
using the
Right , Next
,
and
keys.
A maximum (or minimum) point is detected whenever an amplitude
excursion greater than half of a division occurs. The half of a division
excursion requirement must be satisfied on both sides (left and right) of the
peak (or minimum). The maximum (or minimum) point must be
See Figure 4-4.
I
I
I
All Meet
X Does Not Meet
Peak Search
Peak Search
Criteria
Criteria
Figure 4-4. Peak and Minimum Search Criteria
When the maximum or minimum point is at or near either edge of the
display, the excursion requirement is satisfied by a half of a division excursion
on just one side of the maximum (or minimum). See Figure 4-5.
4-10
Summary of Contents for 8712C
Page 1: ...I I User s Guide HP 8712C and HP 8714C RF Network Analyzers I ...
Page 17: ...I 1 Installing the Analyzer ...
Page 35: ...I 2 Getting Started ...
Page 54: ...I 3 Making Measurements ...
Page 81: ......
Page 119: ...4 Using Instrument Functions ...
Page 182: ...Using Instrument Functions Customizing the Display Figure 4 24 Expanded Display 4 64 ...
Page 223: ...5 Optimizing Measurements ...
Page 244: ...6 Calibrating for Increased Measurement Accuracy ...
Page 280: ...7 Automating Measurements ...
Page 298: ......
Page 362: ...8 Front Rear Panel ...
Page 373: ......
Page 381: ......
Page 386: ... 9 w I Suftkey Reference ...
Page 477: ...I 10 Specifications and Characteristics ...
Page 501: ...11 Safety and Regulatory Information ...
Page 508: ...I 12 preset State and Memory Allocation ...
Page 527: ...Index ...
Page 559: ...Y X 9 91 YYW m m IiH rn 9 91 Z zeroing detectors 6 15 Index 33 ...