I
Using Instrument Functions
Using limit Testing
Delta Frequency
This marker limit test allows you to set marker 1 as a frequency reference
against which marker 2 is limit tested.
1.
2.
3.
4.
5.
This limit test requires that you first use marker 1 to determine the
reference frequency: Press
2
a n d t h e n u s e t h e f r o n t p a n e l
knob or the
keys to place marker 1 at the desired place on the
measurement trace. The frequency of marker 1 at this point becomes the
delta reference’ for this marker limit test.
Press
Limit
Use the front panel knob or the
keys to select Delta
in
the marker limit test table. Turn this limit function on by pressing the
OFF key. Note that the entry in the on/off column of the
table changes to “on.”
Press Edit Limit
Limit and enter the maximum
limit using the front panel keypad and terminating the entry with the
k
e
y
.
Press Edit Limit
Limit and enter the minimum limit
using the front panel keypad and terminating the entry with the
key.
4-43
Summary of Contents for 8712C
Page 1: ...I I User s Guide HP 8712C and HP 8714C RF Network Analyzers I ...
Page 17: ...I 1 Installing the Analyzer ...
Page 35: ...I 2 Getting Started ...
Page 54: ...I 3 Making Measurements ...
Page 81: ......
Page 119: ...4 Using Instrument Functions ...
Page 182: ...Using Instrument Functions Customizing the Display Figure 4 24 Expanded Display 4 64 ...
Page 223: ...5 Optimizing Measurements ...
Page 244: ...6 Calibrating for Increased Measurement Accuracy ...
Page 280: ...7 Automating Measurements ...
Page 298: ......
Page 362: ...8 Front Rear Panel ...
Page 373: ......
Page 381: ......
Page 386: ... 9 w I Suftkey Reference ...
Page 477: ...I 10 Specifications and Characteristics ...
Page 501: ...11 Safety and Regulatory Information ...
Page 508: ...I 12 preset State and Memory Allocation ...
Page 527: ...Index ...
Page 559: ...Y X 9 91 YYW m m IiH rn 9 91 Z zeroing detectors 6 15 Index 33 ...