Using Instrument Functions
Using limit Testing
Sometimes you may only be interested in the level at one particular
frequency. In this case, you may wish to use a single point limit. Using the
setup from the previous examples and building on them, let’s
that
when testing your
it is specified that the insertion loss at
174 MHz must be less than 3
The following example creates a single point
limit at -3
at 174 MHz.
C A U T I O N
Limit tests are only performed on actual data points,
the interpolated
values between. When setting a single point limit, the limit is actually
applied to the closest data point to the frequency of the set limit. See
“Additional Notes on Limit Testing,” later in this chapter for more
information.
1. Press
Add Limit Add
.
Press
(174)
Press Limit
4-37
Summary of Contents for 8712C
Page 1: ...I I User s Guide HP 8712C and HP 8714C RF Network Analyzers I ...
Page 17: ...I 1 Installing the Analyzer ...
Page 35: ...I 2 Getting Started ...
Page 54: ...I 3 Making Measurements ...
Page 81: ......
Page 119: ...4 Using Instrument Functions ...
Page 182: ...Using Instrument Functions Customizing the Display Figure 4 24 Expanded Display 4 64 ...
Page 223: ...5 Optimizing Measurements ...
Page 244: ...6 Calibrating for Increased Measurement Accuracy ...
Page 280: ...7 Automating Measurements ...
Page 298: ......
Page 362: ...8 Front Rear Panel ...
Page 373: ......
Page 381: ......
Page 386: ... 9 w I Suftkey Reference ...
Page 477: ...I 10 Specifications and Characteristics ...
Page 501: ...11 Safety and Regulatory Information ...
Page 508: ...I 12 preset State and Memory Allocation ...
Page 527: ...Index ...
Page 559: ...Y X 9 91 YYW m m IiH rn 9 91 Z zeroing detectors 6 15 Index 33 ...