Automating Measurements
Measurement Setup and Control with Fast Recall
When fast recall is used in conjunction with a switch connected to the
USER TTL IN/OUT rear panel connector, you can cycle through up to seven
instrument states in sequence by activating the switch.
1. Connect a switch to the USER TTL IN/OUT rear panel connector as shown
in Figure 7-7.
SWITCH;
Figure
Connect a Switch to the USER TTL
Connector
2. Make sure the analyzer is configured to use the USER TTL
IN/OUT connector for
sequencing: press
User
3. With fast recall toggled to ON, press the switch several times while
observing the analyzer.
4. Notice that with each press of the switch, the files are highlighted (boxed)
in succession, and that after the last available file has been used, the
sequence starts again at the top of the
menu.
7-43
Summary of Contents for 8712C
Page 1: ...I I User s Guide HP 8712C and HP 8714C RF Network Analyzers I ...
Page 17: ...I 1 Installing the Analyzer ...
Page 35: ...I 2 Getting Started ...
Page 54: ...I 3 Making Measurements ...
Page 81: ......
Page 119: ...4 Using Instrument Functions ...
Page 182: ...Using Instrument Functions Customizing the Display Figure 4 24 Expanded Display 4 64 ...
Page 223: ...5 Optimizing Measurements ...
Page 244: ...6 Calibrating for Increased Measurement Accuracy ...
Page 280: ...7 Automating Measurements ...
Page 298: ......
Page 362: ...8 Front Rear Panel ...
Page 373: ......
Page 381: ......
Page 386: ... 9 w I Suftkey Reference ...
Page 477: ...I 10 Specifications and Characteristics ...
Page 501: ...11 Safety and Regulatory Information ...
Page 508: ...I 12 preset State and Memory Allocation ...
Page 527: ...Index ...
Page 559: ...Y X 9 91 YYW m m IiH rn 9 91 Z zeroing detectors 6 15 Index 33 ...