Optimizing Measurements
Increasing
Sweep Speed
Auto sweep time mode (the preset instrument mode), maintains the fastest
sweep speed possible for any particular measurement settings.
Press
and look at the
label.
When AUTO is all capital letters, it indicates that the analyzer is in auto
sweep time mode. If MAN is all capital letters, the analyzer is in manual
sweep time mode.
necessary, press
to toggle the time mode to
AUTO.
Wide system bandwidth is recommended for some broadband detection
measurements.
Press
to widen the IF’ bandwidth. As the
bandwidth increases, the sweep time decreases and the trace noise increases.
Table
Relationship Between System Bandwidth and Sweep Speed
Bandwidth
Wide
Medium Wide
Medium Narrow
Narrow
Fine
Speed
fastest
faster
fast
slow
slower
I slowest
5-4
Summary of Contents for 8712C
Page 1: ...I I User s Guide HP 8712C and HP 8714C RF Network Analyzers I ...
Page 17: ...I 1 Installing the Analyzer ...
Page 35: ...I 2 Getting Started ...
Page 54: ...I 3 Making Measurements ...
Page 81: ......
Page 119: ...4 Using Instrument Functions ...
Page 182: ...Using Instrument Functions Customizing the Display Figure 4 24 Expanded Display 4 64 ...
Page 223: ...5 Optimizing Measurements ...
Page 244: ...6 Calibrating for Increased Measurement Accuracy ...
Page 280: ...7 Automating Measurements ...
Page 298: ......
Page 362: ...8 Front Rear Panel ...
Page 373: ......
Page 381: ......
Page 386: ... 9 w I Suftkey Reference ...
Page 477: ...I 10 Specifications and Characteristics ...
Page 501: ...11 Safety and Regulatory Information ...
Page 508: ...I 12 preset State and Memory Allocation ...
Page 527: ...Index ...
Page 559: ...Y X 9 91 YYW m m IiH rn 9 91 Z zeroing detectors 6 15 Index 33 ...