Step 3: Verify
Performance
Calibrating for Increased Measurement
Accuracy
Choose an Appropriate Calibration Method
Once a measurement calibration has been generated with a user-defined
calibration kit, its performance should be checked before making device
measurements. lb check the accuracy that can be obtained using the new
calibration kit, a device with a well-defined frequency response (preferably
any of the standards used) should be measured. The
device must not be one of the calibration standards: measurement of one of
these standards is merely a measure of repeatability.
‘lb achieve more complete
of a particular measurement
calibration, accurately known verification standards with a diverse magnitude
and phase response should be used. NBS traceable or HP standards are
recommended to achieve
measurement accuracy.
C A U T I O N
The published
for your analyzer system include accuracy
enhancement with compatible calibration kits. Measurement calibrations
made with
or modified calibration kits are not subject to those
analyzer
6-24
Summary of Contents for 8712C
Page 1: ...I I User s Guide HP 8712C and HP 8714C RF Network Analyzers I ...
Page 17: ...I 1 Installing the Analyzer ...
Page 35: ...I 2 Getting Started ...
Page 54: ...I 3 Making Measurements ...
Page 81: ......
Page 119: ...4 Using Instrument Functions ...
Page 182: ...Using Instrument Functions Customizing the Display Figure 4 24 Expanded Display 4 64 ...
Page 223: ...5 Optimizing Measurements ...
Page 244: ...6 Calibrating for Increased Measurement Accuracy ...
Page 280: ...7 Automating Measurements ...
Page 298: ......
Page 362: ...8 Front Rear Panel ...
Page 373: ......
Page 381: ......
Page 386: ... 9 w I Suftkey Reference ...
Page 477: ...I 10 Specifications and Characteristics ...
Page 501: ...11 Safety and Regulatory Information ...
Page 508: ...I 12 preset State and Memory Allocation ...
Page 527: ...Index ...
Page 559: ...Y X 9 91 YYW m m IiH rn 9 91 Z zeroing detectors 6 15 Index 33 ...