I
Making Measurements
Using the BEGIN Key to Make Measurements
N O T E
If the new measurement selected is a broadband measurement such as power, conversion loss, or AM
the start frequency is limited to at least 10 MHz. Therefore, if your customized setup contains
a start frequency below 10 MHz and you choose power, conversion loss, or AM delay, the start
frequency will be changed to 10 MHz. The stop frequency will remain unchanged, unless it was sat to
below 10 MHz.
The (BEGIN) Kay and
The (BEGIN) key is designed to work when measurement channel 1 is active.
Measurement Channels
However, it does change the measurement mode of measurement channel 2
as well.
If measurement channel 2 is active when the
key is used to select a
new
type, measurement channel 2 is turned off, and measurement
channel 1 is made active.
If measurement channel 2 is active when the
key is used to select a
new
measurement
type, measurement channel 2 will be left on
and
active.
However, the analyzer then proceeds to setup channel 1 for the requested
measurement type, even though channel 2 is the active channel.
3-14
Summary of Contents for 8712C
Page 1: ...I I User s Guide HP 8712C and HP 8714C RF Network Analyzers I ...
Page 17: ...I 1 Installing the Analyzer ...
Page 35: ...I 2 Getting Started ...
Page 54: ...I 3 Making Measurements ...
Page 81: ......
Page 119: ...4 Using Instrument Functions ...
Page 182: ...Using Instrument Functions Customizing the Display Figure 4 24 Expanded Display 4 64 ...
Page 223: ...5 Optimizing Measurements ...
Page 244: ...6 Calibrating for Increased Measurement Accuracy ...
Page 280: ...7 Automating Measurements ...
Page 298: ......
Page 362: ...8 Front Rear Panel ...
Page 373: ......
Page 381: ......
Page 386: ... 9 w I Suftkey Reference ...
Page 477: ...I 10 Specifications and Characteristics ...
Page 501: ...11 Safety and Regulatory Information ...
Page 508: ...I 12 preset State and Memory Allocation ...
Page 527: ...Index ...
Page 559: ...Y X 9 91 YYW m m IiH rn 9 91 Z zeroing detectors 6 15 Index 33 ...