I
Optimizing Measurements
Increasing Network Analyzer Dynamic Range
Changing Measurement
In averaging mode, the analyzer measures each frequency point once per
Averaging
sweep and averages the current and previous trace up to the averaging factor
by the user. The instrument computes each data point based on an
exponential average of consecutive sweeps weighted by the user-specified
averaging factor.
As the averaging factor is increased:
signal-to-noise ratio increases
l
time for each individual sweep remains the same, but
l
total time to update the trace increases.
Averaging is better than system bandwidth reduction at minimizing very
low
frequency noise.
Press
.
2. Enter a value followed by
Press
ON
.
N O T E
When averaging is on, there is an indicator in the lower right corner of the analyzer’s display The
display is in the form
Avg
where is the averaging factor for measurement channel 1, and
is the averaging factor for measurement channel 2. If only one measurement channel is on, the
averaging indicator only displays that channel’s averaging factor.
5-12
Summary of Contents for 8712C
Page 1: ...I I User s Guide HP 8712C and HP 8714C RF Network Analyzers I ...
Page 17: ...I 1 Installing the Analyzer ...
Page 35: ...I 2 Getting Started ...
Page 54: ...I 3 Making Measurements ...
Page 81: ......
Page 119: ...4 Using Instrument Functions ...
Page 182: ...Using Instrument Functions Customizing the Display Figure 4 24 Expanded Display 4 64 ...
Page 223: ...5 Optimizing Measurements ...
Page 244: ...6 Calibrating for Increased Measurement Accuracy ...
Page 280: ...7 Automating Measurements ...
Page 298: ......
Page 362: ...8 Front Rear Panel ...
Page 373: ......
Page 381: ......
Page 386: ... 9 w I Suftkey Reference ...
Page 477: ...I 10 Specifications and Characteristics ...
Page 501: ...11 Safety and Regulatory Information ...
Page 508: ...I 12 preset State and Memory Allocation ...
Page 527: ...Index ...
Page 559: ...Y X 9 91 YYW m m IiH rn 9 91 Z zeroing detectors 6 15 Index 33 ...