Making Measurements
Measuring Devices with Your Network Analyzer
Refer to Figure 3-3 for the following discussion. The network analyzer
receiver has two signal detection modes:
l
broadband detection mode
l
narrowband detection mode
There are two internal broadband detector inputs: B* and
External
broadband detectors can also be used when connected to the X and Y
ports on the rear panel of the analyzer. When the network analyzer is in
the broadband detection mode, it measures the total power of all signals
present at these measurement ports, independent of signal frequency. This
enables the characterization of frequency translation devices such as mixers,
receivers, and tuners, where the RF input and output frequencies are not the
same. Figure 3-3 labels the transmitted signal for broadband detection input
as
and the reference signal as R*.
When the network analyzer is in the narrowband detection mode, the
receiver is tuned to the source frequency. This technique provides greater
dynamic range by decreasing the receiver’s bandwidth. Figure 3-3 shows the
transmitted signal for narrowband detection input as B, and the reference
signal as R.
3-6
Summary of Contents for 8712C
Page 1: ...I I User s Guide HP 8712C and HP 8714C RF Network Analyzers I ...
Page 17: ...I 1 Installing the Analyzer ...
Page 35: ...I 2 Getting Started ...
Page 54: ...I 3 Making Measurements ...
Page 81: ......
Page 119: ...4 Using Instrument Functions ...
Page 182: ...Using Instrument Functions Customizing the Display Figure 4 24 Expanded Display 4 64 ...
Page 223: ...5 Optimizing Measurements ...
Page 244: ...6 Calibrating for Increased Measurement Accuracy ...
Page 280: ...7 Automating Measurements ...
Page 298: ......
Page 362: ...8 Front Rear Panel ...
Page 373: ......
Page 381: ......
Page 386: ... 9 w I Suftkey Reference ...
Page 477: ...I 10 Specifications and Characteristics ...
Page 501: ...11 Safety and Regulatory Information ...
Page 508: ...I 12 preset State and Memory Allocation ...
Page 527: ...Index ...
Page 559: ...Y X 9 91 YYW m m IiH rn 9 91 Z zeroing detectors 6 15 Index 33 ...