Using Instrument Functions
Using limit Testing
Peak-to-Peak Ripple
1. This limit test requires that you first define a segment on the measurement
trace using markers 1 and 2 (or markers 3 and 4 for measurement
channel 2). Then press [MARKER)
to enable the statistics marker search. See “lb Use Marker
Statistics” earlier in this chapter for more information on using marker
statistics.
2. Press
.
3. Use the front panel knob or the
keys to select Statistic: p-p in
the marker limit test table. Turn this limit function on by pressing the
OFF key. Note that the entry in the on/off column of the
table changes to “on.”
4. Press
and enter the maximum
limit using the front panel keypad and terminating the entry with the
k e y .
Press
Limit
and enter the minimum limit
using the front panel keypad and terminating the entry with the
key.
N O T E
Note that there are no visible limit lines or indicators on the display with this limit test function.
Therefore the
function has no effect when using marker limits.
4-40
Summary of Contents for 8712C
Page 1: ...I I User s Guide HP 8712C and HP 8714C RF Network Analyzers I ...
Page 17: ...I 1 Installing the Analyzer ...
Page 35: ...I 2 Getting Started ...
Page 54: ...I 3 Making Measurements ...
Page 81: ......
Page 119: ...4 Using Instrument Functions ...
Page 182: ...Using Instrument Functions Customizing the Display Figure 4 24 Expanded Display 4 64 ...
Page 223: ...5 Optimizing Measurements ...
Page 244: ...6 Calibrating for Increased Measurement Accuracy ...
Page 280: ...7 Automating Measurements ...
Page 298: ......
Page 362: ...8 Front Rear Panel ...
Page 373: ......
Page 381: ......
Page 386: ... 9 w I Suftkey Reference ...
Page 477: ...I 10 Specifications and Characteristics ...
Page 501: ...11 Safety and Regulatory Information ...
Page 508: ...I 12 preset State and Memory Allocation ...
Page 527: ...Index ...
Page 559: ...Y X 9 91 YYW m m IiH rn 9 91 Z zeroing detectors 6 15 Index 33 ...