Using Instrument Functions
Using limit Testing
Stimulus and Amplitude
Values
C A U T I O N
Calculating the
Frequency of a Data
Point
In frequency sweep mode, the stimulus values are interpreted as frequencies;
in power sweep mode, the stimulus values are interpreted as output power
levels.
The values entered for stimulus and amplitude are unitless. If the
measurement format is changed after limit lines are set, the amplitude values
do not automatically change. Therefore, you should select the format you
want to use before entering your limits.
The frequency of each data point is calculated using the following formula:
Freq @t-number) =
(pt-number
1)
x (stop-freq start-freq)
number-of -points
1
+ start-freq
where
pt-number
ranges from 1 (for the leftmost display point) to
(the rightmost display point).
limit Testing and
Measurement Points
Limit testing is only performed on actual data points,
not
interpolated values
between them. The limit lines or points that you enter are converted
to values at each measurement point. Most of the time, this has little or
no effect on the validity of the limit testing. However, note the following
examples of problems that could arise when using limit lines:
4-47
Summary of Contents for 8712C
Page 1: ...I I User s Guide HP 8712C and HP 8714C RF Network Analyzers I ...
Page 17: ...I 1 Installing the Analyzer ...
Page 35: ...I 2 Getting Started ...
Page 54: ...I 3 Making Measurements ...
Page 81: ......
Page 119: ...4 Using Instrument Functions ...
Page 182: ...Using Instrument Functions Customizing the Display Figure 4 24 Expanded Display 4 64 ...
Page 223: ...5 Optimizing Measurements ...
Page 244: ...6 Calibrating for Increased Measurement Accuracy ...
Page 280: ...7 Automating Measurements ...
Page 298: ......
Page 362: ...8 Front Rear Panel ...
Page 373: ......
Page 381: ......
Page 386: ... 9 w I Suftkey Reference ...
Page 477: ...I 10 Specifications and Characteristics ...
Page 501: ...11 Safety and Regulatory Information ...
Page 508: ...I 12 preset State and Memory Allocation ...
Page 527: ...Index ...
Page 559: ...Y X 9 91 YYW m m IiH rn 9 91 Z zeroing detectors 6 15 Index 33 ...