Installing the Analyzer
Step 2. Meet Electrical and Environmental Requirements
2. Ensure the available ac power source meets the following requirements:
115
90 to 132
to 66 Hzl
I
230
V
198 to 254
to 66 Hzl
If the ac line voltage does not fall within these ranges, an autotransformer
that provides third wire continuity to ground should be used.
3. Ensure the operating environment meets the following requirements for
safety:
l
indoor use
l
altitude up to 15,000 feet (4,572 meters)
temperature 0
to 55
l
maximum relative humidity 5 to 95 percent relative at
(non-condensing)
l
this product is designed for use in INSTALLATION CATEGORY II and
POLLUTION DEGREE 2
C A U T I O N
This product is designed for use in Installation Category II and Pollution
Degree 2.
N O T E
The above requirements are for safety only Separate conditions that must be met for specified
performance are noted in Chapter 10.
Summary of Contents for 8712C
Page 1: ...I I User s Guide HP 8712C and HP 8714C RF Network Analyzers I ...
Page 17: ...I 1 Installing the Analyzer ...
Page 35: ...I 2 Getting Started ...
Page 54: ...I 3 Making Measurements ...
Page 81: ......
Page 119: ...4 Using Instrument Functions ...
Page 182: ...Using Instrument Functions Customizing the Display Figure 4 24 Expanded Display 4 64 ...
Page 223: ...5 Optimizing Measurements ...
Page 244: ...6 Calibrating for Increased Measurement Accuracy ...
Page 280: ...7 Automating Measurements ...
Page 298: ......
Page 362: ...8 Front Rear Panel ...
Page 373: ......
Page 381: ......
Page 386: ... 9 w I Suftkey Reference ...
Page 477: ...I 10 Specifications and Characteristics ...
Page 501: ...11 Safety and Regulatory Information ...
Page 508: ...I 12 preset State and Memory Allocation ...
Page 527: ...Index ...
Page 559: ...Y X 9 91 YYW m m IiH rn 9 91 Z zeroing detectors 6 15 Index 33 ...