I
Making Measurements
Measuring Devices with Your Network Analyzer
Analyzers with a system characteristic impedance of 50 or 75 ohms, may
be switched to the alternate impedance. If using minimum loss pads for
impedance conversions, the alternate impedance should be selected so that
the measurement results are displayed relative to the conversion impedance.
For example, if you have a 50 ohm instrument and are making 75 ohm
measurements, you may be using a 50 to 75 ohm minimum loss pad.
Measurement results can be reported relative to 75 ohms, not 50 ohms, if the
alternate system impedance is selected. This may include marker readouts,
Smith chart results, or SRL impedance computations (Option 100).
change the system impedance, press the following keys on the analyzer:
(CAL)
3-10
The built-in cal kit selections will be converted to the selected system
impedance.
Summary of Contents for 8712C
Page 1: ...I I User s Guide HP 8712C and HP 8714C RF Network Analyzers I ...
Page 17: ...I 1 Installing the Analyzer ...
Page 35: ...I 2 Getting Started ...
Page 54: ...I 3 Making Measurements ...
Page 81: ......
Page 119: ...4 Using Instrument Functions ...
Page 182: ...Using Instrument Functions Customizing the Display Figure 4 24 Expanded Display 4 64 ...
Page 223: ...5 Optimizing Measurements ...
Page 244: ...6 Calibrating for Increased Measurement Accuracy ...
Page 280: ...7 Automating Measurements ...
Page 298: ......
Page 362: ...8 Front Rear Panel ...
Page 373: ......
Page 381: ......
Page 386: ... 9 w I Suftkey Reference ...
Page 477: ...I 10 Specifications and Characteristics ...
Page 501: ...11 Safety and Regulatory Information ...
Page 508: ...I 12 preset State and Memory Allocation ...
Page 527: ...Index ...
Page 559: ...Y X 9 91 YYW m m IiH rn 9 91 Z zeroing detectors 6 15 Index 33 ...