Optimizing Measurements
Increasing Sweep Speed
Alternate sweep is turned off when the analyzer is preset, but is
automatically activated with some dual channel measurements. The alternate
sweep feature sweeps and measures one channel at a time. By disengaging
this feature, you increase the sweep speed by 50 percent.
Press
and look at the
2. If ON is in capital letters, toggle the
to disengage the alternate
sweep feature (the
should appear
When markers are on, time is required to update the marker readouts.
Turning off markers can reduce sweep cycle time by up to 30 ms. When the
marker
menu is active, it too must be updated, adding up to 10 ms to
the sweep cycle time.
All
When marker tracking is on, the analyzer performs a search after each sweep.
Simple searches like
can take 10 to 20 ms per sweep. Lengthy
calculations like
can add over 100 ms to each sweep.
Press
Marker
and look at the
2. If the word “ON” is in capital letters, toggle the
to the OFF’
position.
5-7
Summary of Contents for 8712C
Page 1: ...I I User s Guide HP 8712C and HP 8714C RF Network Analyzers I ...
Page 17: ...I 1 Installing the Analyzer ...
Page 35: ...I 2 Getting Started ...
Page 54: ...I 3 Making Measurements ...
Page 81: ......
Page 119: ...4 Using Instrument Functions ...
Page 182: ...Using Instrument Functions Customizing the Display Figure 4 24 Expanded Display 4 64 ...
Page 223: ...5 Optimizing Measurements ...
Page 244: ...6 Calibrating for Increased Measurement Accuracy ...
Page 280: ...7 Automating Measurements ...
Page 298: ......
Page 362: ...8 Front Rear Panel ...
Page 373: ......
Page 381: ......
Page 386: ... 9 w I Suftkey Reference ...
Page 477: ...I 10 Specifications and Characteristics ...
Page 501: ...11 Safety and Regulatory Information ...
Page 508: ...I 12 preset State and Memory Allocation ...
Page 527: ...Index ...
Page 559: ...Y X 9 91 YYW m m IiH rn 9 91 Z zeroing detectors 6 15 Index 33 ...