Installing the Analyzer
Step 4. Configure the Analyzer
HP-IB Connections
An HP-IB system may be connected in any
as long as the
following rules are observed:
l
The total number of devices is less than or equal to 15.
l
The total length of all the cables used is less than or equal to 2 meters
times the number of devices connected together up to an absolute
maximum of 20 meters. For example, the maximum cable length is
4 meters if only 2 devices are involved. The length between adjacent
devices is not critical as long as the overall restriction is met.
See Figure
for different connection configurations.
S T A R
L I N E A R
NETWORK
NETWORK
ANALYZER
ANALYZER
Figure
HP-IB Connection Configurations
1-13
Summary of Contents for 8712C
Page 1: ...I I User s Guide HP 8712C and HP 8714C RF Network Analyzers I ...
Page 17: ...I 1 Installing the Analyzer ...
Page 35: ...I 2 Getting Started ...
Page 54: ...I 3 Making Measurements ...
Page 81: ......
Page 119: ...4 Using Instrument Functions ...
Page 182: ...Using Instrument Functions Customizing the Display Figure 4 24 Expanded Display 4 64 ...
Page 223: ...5 Optimizing Measurements ...
Page 244: ...6 Calibrating for Increased Measurement Accuracy ...
Page 280: ...7 Automating Measurements ...
Page 298: ......
Page 362: ...8 Front Rear Panel ...
Page 373: ......
Page 381: ......
Page 386: ... 9 w I Suftkey Reference ...
Page 477: ...I 10 Specifications and Characteristics ...
Page 501: ...11 Safety and Regulatory Information ...
Page 508: ...I 12 preset State and Memory Allocation ...
Page 527: ...Index ...
Page 559: ...Y X 9 91 YYW m m IiH rn 9 91 Z zeroing detectors 6 15 Index 33 ...