I
Making Measurements
Measuring Transmission Response
Your analyzer can provide highly accurate measurements without performing
any additional user-calibrations if certain conditions are met. This example
describes how to perform an enhanced transmission response calibration.
When you perform an enhanced transmission response calibration, the
analyzer performs correction at the selected number of data points across
the selected frequency band. Interpolation recalculates the error correction
array for reduced frequency spans. If the frequency span is increased the
calibration is invalidated, and the default response calibration is automatically
restored.
Chapter 6 provides detail about when additional calibration is necessary, and
information about other calibrations available for transmission measurements.
If you wish to calibrate your instrument for a transmission response
measurement, perform the following steps:
Press
.
2. The instrument prompts you to connect four standards-open, short, load,
and through cable-as shown below.
NETWORK ANALYZER
N E T W O R K A N A L Y Z E R
R F O U T
R F O U T
RF IN
OPEN
SHORT
LOAD
Open, Short, load Connections
Through
3. Press
after connecting each standard.
3-19
Summary of Contents for 8712C
Page 1: ...I I User s Guide HP 8712C and HP 8714C RF Network Analyzers I ...
Page 17: ...I 1 Installing the Analyzer ...
Page 35: ...I 2 Getting Started ...
Page 54: ...I 3 Making Measurements ...
Page 81: ......
Page 119: ...4 Using Instrument Functions ...
Page 182: ...Using Instrument Functions Customizing the Display Figure 4 24 Expanded Display 4 64 ...
Page 223: ...5 Optimizing Measurements ...
Page 244: ...6 Calibrating for Increased Measurement Accuracy ...
Page 280: ...7 Automating Measurements ...
Page 298: ......
Page 362: ...8 Front Rear Panel ...
Page 373: ......
Page 381: ......
Page 386: ... 9 w I Suftkey Reference ...
Page 477: ...I 10 Specifications and Characteristics ...
Page 501: ...11 Safety and Regulatory Information ...
Page 508: ...I 12 preset State and Memory Allocation ...
Page 527: ...Index ...
Page 559: ...Y X 9 91 YYW m m IiH rn 9 91 Z zeroing detectors 6 15 Index 33 ...