I
I
Increasing Sweep Speed
You can increase the analyzer sweep speed by avoiding the use of some
features that require computational time for implementation and updating,
such as bandwidth marker tracking.
You can also increase the sweep speed by making adjustments to the
measurement settings. Listed below are some of the things that can be done
to increase sweep speed.
l
increase the start frequency
l
sweep time in AUTO mode
l
widen the system bandwidth
l
reduce the amount of averaging
l
reduce the number of measurement points
l
only view a single channel
turn off alternate sweep
l
turn off markers and marker tracking
l
turn off spur avoidance
l
minimize frequency span to avoid bandcrossings (HP
only)
Since the analyzer sweeps frequencies below approximately 20 MHz at a
slower rate, you can increase the start frequency to speed up the sweep.
Press
.
2. Enter the highest start frequency possible for your measurement.
5-3
Summary of Contents for 8712C
Page 1: ...I I User s Guide HP 8712C and HP 8714C RF Network Analyzers I ...
Page 17: ...I 1 Installing the Analyzer ...
Page 35: ...I 2 Getting Started ...
Page 54: ...I 3 Making Measurements ...
Page 81: ......
Page 119: ...4 Using Instrument Functions ...
Page 182: ...Using Instrument Functions Customizing the Display Figure 4 24 Expanded Display 4 64 ...
Page 223: ...5 Optimizing Measurements ...
Page 244: ...6 Calibrating for Increased Measurement Accuracy ...
Page 280: ...7 Automating Measurements ...
Page 298: ......
Page 362: ...8 Front Rear Panel ...
Page 373: ......
Page 381: ......
Page 386: ... 9 w I Suftkey Reference ...
Page 477: ...I 10 Specifications and Characteristics ...
Page 501: ...11 Safety and Regulatory Information ...
Page 508: ...I 12 preset State and Memory Allocation ...
Page 527: ...Index ...
Page 559: ...Y X 9 91 YYW m m IiH rn 9 91 Z zeroing detectors 6 15 Index 33 ...