I
Using Instrument Functions
Using Markers
To Use Marker Flatness
The marker flatness search function measures a user-defined segment of the
measurement trace and calculates the following:
frequency span
gain
slope
flatness
The analyzer calculates flatness by drawing a straight line between the
markers. A maximum vertical deviation from this line is computed for each
measurement point. Flatness is the magnitude difference of the maximum
and minimum calculated deviations from the straight line.
Limit testing may be performed on the flatness parameter. See
Use
Marker Limit Functions,’ later in this chapter for information.
1. On measurement channel 1 press
and set markers 1 and 2
to define the beginning and end of the trace segment that you want to
measure. (When using measurement channel 2, use markers 3 and 4 to
define the trace segment.)
3. Figure 4-13 shows a
trace segment. Notice the marker readout in
the upper right corner of the display.
4-23
Summary of Contents for 8712C
Page 1: ...I I User s Guide HP 8712C and HP 8714C RF Network Analyzers I ...
Page 17: ...I 1 Installing the Analyzer ...
Page 35: ...I 2 Getting Started ...
Page 54: ...I 3 Making Measurements ...
Page 81: ......
Page 119: ...4 Using Instrument Functions ...
Page 182: ...Using Instrument Functions Customizing the Display Figure 4 24 Expanded Display 4 64 ...
Page 223: ...5 Optimizing Measurements ...
Page 244: ...6 Calibrating for Increased Measurement Accuracy ...
Page 280: ...7 Automating Measurements ...
Page 298: ......
Page 362: ...8 Front Rear Panel ...
Page 373: ......
Page 381: ......
Page 386: ... 9 w I Suftkey Reference ...
Page 477: ...I 10 Specifications and Characteristics ...
Page 501: ...11 Safety and Regulatory Information ...
Page 508: ...I 12 preset State and Memory Allocation ...
Page 527: ...Index ...
Page 559: ...Y X 9 91 YYW m m IiH rn 9 91 Z zeroing detectors 6 15 Index 33 ...