Using Instrument Functions
Using limit Testing
Example 2.
In this example, the analyzer has been set up with the following parameters:
Start frequency = 90 MHz
Stop frequency 210 MHz
Number of points = 11
Maximum limit line begin frequency 90 MHz
Maximum limit line end frequency = 200 MHz
Refer to the illustration below for the discussion. Here, the data trace passes
through the limit line, and is above it, but this test passes because the last
point
is at 198 MHz. The last point (210 MHz) does not cause the limit
test to fail because it is past the limit line stop frequency of 200 MHz. But
there is an area on the data trace between 198 MHz and 200 MHz that is
above the displayed limit line.
E N D L I M I T
a t 2 0 0 M H z
M A X I M U M
DATA
L I M I T L I N E
I N Q U E S T I O N
N E X T
I
D A T A P O I N T
a t 2 1 0 M H z
l ’
I
I
A
.-.
9 0
1 0 2 1 1 4 1 2 6 1 3 8 1 5 0 1 6 2 1 7 4 1 8 6 1 9 8
D A T A P O I N T S
Figure 4-18. limit lines Example 2
4 - 4 9
Summary of Contents for 8712C
Page 1: ...I I User s Guide HP 8712C and HP 8714C RF Network Analyzers I ...
Page 17: ...I 1 Installing the Analyzer ...
Page 35: ...I 2 Getting Started ...
Page 54: ...I 3 Making Measurements ...
Page 81: ......
Page 119: ...4 Using Instrument Functions ...
Page 182: ...Using Instrument Functions Customizing the Display Figure 4 24 Expanded Display 4 64 ...
Page 223: ...5 Optimizing Measurements ...
Page 244: ...6 Calibrating for Increased Measurement Accuracy ...
Page 280: ...7 Automating Measurements ...
Page 298: ......
Page 362: ...8 Front Rear Panel ...
Page 373: ......
Page 381: ......
Page 386: ... 9 w I Suftkey Reference ...
Page 477: ...I 10 Specifications and Characteristics ...
Page 501: ...11 Safety and Regulatory Information ...
Page 508: ...I 12 preset State and Memory Allocation ...
Page 527: ...Index ...
Page 559: ...Y X 9 91 YYW m m IiH rn 9 91 Z zeroing detectors 6 15 Index 33 ...