Automating Measurements
Automated Measurement Setup and Control
Tip
If several measurement setups are similar differing only by a few instrument parameters the
fastest way to switch between the states is for the control program to send the
commands that
modify those parameters.
For example, if center frequency and source power are the only parameters
that change in consecutive measurement setups, send the SCPI commands
and SOURl:POW to change these parameters, leaving all
other instrument settings unchanged. This will be faster than either recalling
an instrument state or sending a learn string.
7-49