I
I
Calibrating for Increased Measurement
Accuracy
Check
Calibration
‘lb perform a calibration check:
1. Be sure that the analyzer is in the measurement mode and has
implemented the particular calibration that you want to check.
Press
Do Cal Check,
3. Follow the
prompts to connect the calibration standards.
(You should connect the standards at your calibration reference plane,
not necessarily at the front panel of the analyzer. For example, your
calibration reference plane may be at the end of cables, adapters, or test
connected to the analyzer. See “The Calibration Reference Plane”
for more information.)
C A U T I O N
decrease the likelihood that a faulty calibration standard is being used,
it is recommended that you use a
set of calibration standards for
the calibration check than you used to perform the initial calibration. The
different set of calibration standards must be of the same type as the original
set.
4. Once you have properly connected all the required standards as prompted,
the analyzer computes the residual errors for the measurement.
Now press the
View
key to view the residual errors for the
type of calibration you did a calibration check on. When you view an
error term plot, the analyzer stops sweeping and displays the plot on the
analyzer’s screen. lb return to measurement mode and autoscale the
display, press
.
6. Refer to “Error Term Descriptions and Typical Values” for descriptions and
typical plots of the various error terms.
The following table lists the typical error term values.
6-28
Summary of Contents for 8712C
Page 1: ...I I User s Guide HP 8712C and HP 8714C RF Network Analyzers I ...
Page 17: ...I 1 Installing the Analyzer ...
Page 35: ...I 2 Getting Started ...
Page 54: ...I 3 Making Measurements ...
Page 81: ......
Page 119: ...4 Using Instrument Functions ...
Page 182: ...Using Instrument Functions Customizing the Display Figure 4 24 Expanded Display 4 64 ...
Page 223: ...5 Optimizing Measurements ...
Page 244: ...6 Calibrating for Increased Measurement Accuracy ...
Page 280: ...7 Automating Measurements ...
Page 298: ......
Page 362: ...8 Front Rear Panel ...
Page 373: ......
Page 381: ......
Page 386: ... 9 w I Suftkey Reference ...
Page 477: ...I 10 Specifications and Characteristics ...
Page 501: ...11 Safety and Regulatory Information ...
Page 508: ...I 12 preset State and Memory Allocation ...
Page 527: ...Index ...
Page 559: ...Y X 9 91 YYW m m IiH rn 9 91 Z zeroing detectors 6 15 Index 33 ...