![Hioki MR8880-20 Instruction Manual Download Page 284](http://html1.mh-extra.com/html/hioki/mr8880-20/mr8880-20_instruction-manual_2137332284.webp)
Appendix 8 Frequently Asked Questions
A
22
How do I acquire data only
when the external signal is in
the HIGH period?
This is only possible with the Real-
time Function.
Input an external signal to the
EXT.TRIG terminal, and configure
the settings as follows.
[Ext trig]-[Start:
↑
(Rising)], [Stop:
↓
(Falling)]
This will enable data to be acquired
only when the external signal is in
the HIGH period.
"Applying Trigger with
External Input Signal
(External Trigger)" (p.
127)
Can I use two instruments and
synchronize the measure-
ment?
The start of measurement can be
synchronized using an external trig-
ger. It is not possible to synchronize
sampling itself.
"12.6 Synchronous
Measurements with
Multiple Instruments"
(p. 226)
Measuring
Question
Remedial Action
Reference page
What should I do about zero
offset when the input is short-
ed?
Execute zero adjustment to com-
pensate for offset when the inputs
are shorted.
"2.8 Compensating for
Input Circuit Offset (Ze-
ro Adjustment)" (p. 49)
Even though a signal is only in-
put on CH1, why do similar
waveforms appear on unused
channels?
An open-circuit input terminal can be
affected by signals on other chan-
nels. By closing the input terminal
circuit, normal measurement is pos-
sible.
If you want to turn off the waveforms,
set the display of unused channels
to Off or short between the + and –
terminals.
"Setting Waveform Dis-
play Colors & Turning
On/Off Display" (p. 96)
How to display numerical data
values?
Press the
MONITOR
key to confirm
the input channels.
Cursors A and B can be used to read
the numerical values of specified
places.
"Checking Input State
(Monitor)" (p. 55)
"Displaying Cursor Val-
ues" (p. 138)
How can I apply markers while
measuring for easy searching
later?
An event mark can be inserted.
"5.2 Marking Wave-
forms and Searching
Marks (Search Func-
tion)" (p. 141)
Question
Remedial Action
Reference page
Summary of Contents for MR8880-20
Page 2: ......
Page 12: ...Procedure 2 ...
Page 40: ...1 3 Screen Configuration and Display 30 ...
Page 76: ...3 4 Measurement Using Setting Wizard PRESETS key 66 ...
Page 174: ...6 4 Data Management 164 ...
Page 186: ...7 7 Performing a Printer Check 176 ...
Page 196: ...8 4 Numerical Value Calculation Expressions 186 ...
Page 216: ...10 4 Making System Settings 206 ...
Page 222: ...11 3 Communication Using Commands 212 4Click Next to start install ing Installing Click ...
Page 226: ...11 3 Communication Using Commands 216 ...
Page 254: ...13 6 Setting Wizard Function PRESETS 244 ...
Page 292: ...Appendix 10 Installing Waveform Viewer A30 ...